ID SE24
Title Combination of Liquid Chromatography−Fourier Transform Ion Cyclotron Resonance-Mass Spectrometry with 13C‑Labeling for Chemical Assignment of Sulfur-Containing Metabolites in Onion Bulbs
Description We provide a chemical assignment strategy for S-metabolites using 12C- and 13C-based MS and tandem mass spectrometry (MS/MS) spectra recorded by Fourier transform ion cyclotron resonance-mass spectrometry (FTICR-MS) coupled with liquid chromatography (LC) as a case study of the metabolite of onion bulbs. We performed comprehensive peak picking using the theoretical mass difference (1.99579 Da) between 32S-containing monoisotopic ions and their 34S-substituted counterparts for 67 S-containing monoisotopic ions. From these, we identified the specific elemental composition of 22 ions. Finally, we determined that six ions were derived from S-alk(en)ylcysteine sulfoxides and glutathione derivatives.
Authors Ryo Nakabayashi, Yuji Sawada, Yutaka Yamada, Makoto Suzuki, Masami Yokota Hirai, Tetsuya Sakurai, and Kazuki Saito
Reference Nakabayashi R et al. (2013) Analytical Chemistry 85: 1310-1315
Comment The metadata is prepared by the Metabolonote administrator (Sakurai N).


Link icon database.png Link icon dropmet.png

The raw data files are available at DROP Met web site in PRIMe database of RIKEN.

Link icon database.png Link icon pgdbj.png

The web resources and information related to the species used in this study are available at Plant Genome Database Japan (PGDBj).

Link icon article.png

Analytical Method Details Information

ID MS1
Title LC-FTICR-MS Analysis (Positive)
Instrument LC, Agilent 1200 series (Agilent); MS, Solarix 7.0T (Bruker Daltonics)
Instrument Type LC-FTICR-MS
Ionization ESI
Ion Mode Positive
Description Extraction of Metabolites:

The freeze-dried samples were extracted with 50 uL of 80% MeOH containing 2.5 uM lidocaine per mg dry weight using a mixer mill (MM300, Retsch) with zirconia beads for 10 min at 20 Hz and 4C. After centrifugation for 10 min, the supernatant was filtered using and HLB uElution plate (Waters).


LC-FTICR-MS Analysis:

A liquid sample (1 uL) was analyzed using LC-FTICR-MS (LC, Agilent 1200 series; MS, Bruker Daltonics solarix 7.0 T). Analytical conditions were as follows. LC: column, Xselect CSH C18 (3.5 um, 2.1 mm x 150 mm, Waters); solvent system, solvent A (water with 0.1% formic acid) and solvent B (acetonitrile with 0.1% formic acid); gradient program, 99.5% A/0.5% B at 0 min, 0.5% A/99.5% B at 30.0 min, 0.5% A/99.5% B at 45.min, 99.5% A/0.5% B at 45.1 min, and 99.5% A/0.5% B at 60.0 min; flow rate, 0.3 mL/min; column temperature, 35 C; wavelength, 200-600 nm; MS detection (resolution 260000 at m/z 400); polarity, positive; calibrant, low concentration TuningMix (Agilent); lock mass target, diethylhexylphthalate (C24H38O4, [M+H]+ m/z 391.28429); MS/MS detection (resolution 35000 at m/z 400).