SE53:/MS03
Sample Set Information
ID | SE53 |
---|---|
Title | Covering chemical diversity of genetically-modified tomatoes using metabolomics for objective substantial equivalence assessment |
Description | We propose using multiple analytical platforms for the direct acquisition of an interpretable data set of estimable chemical diversity. As an example, we report an application of our multi-platform approach that assesses the substantial equivalence of tomatoes over-expressing the taste-modifying protein miraculin. In combination, the chosen platforms detected compounds that represent 86% of the estimated chemical diversity of the metabolites listed in the LycoCyc database. Following a proof-of-safety approach, we show that w92% had an acceptable range of variation while simultaneously indicating a reproducible transformation-related metabolic signature. We conclude that multi-platform metabolomics is an approach that is both sensitive and robust and that it constitutes a good starting point for characterizing genetically modified organisms. |
Authors | Miyako Kusano, Henning Redestig, Tadayoshi Hirai, Akira Oikawa, Fumio Matsuda, Atsushi Fukushima, Masanori Arita, Shin Watanabe, Megumu Yano, Kyoko Hiwasa-Tanase, Hiroshi Ezura, Kazuki Saito |
Reference | Kusano M et al. (2011) PLOS ONE 6: e16989 |
Comment |
The raw data files are available at DROP Met web site in PRIMe database of RIKEN.
Analytical Method Details Information
ID | MS03 |
---|---|
Title | CE-TOF/MS |
Instrument | Agilent CE capillary electrophoresis system,Agilent G3250AA LC/MSD TOF system,Agilent 1100 series binary HPLC pump,G1603A Agilent CE-MS adapter and G1607A Agilent CE-ESI-MS sprayer kit |
Instrument Type | |
Ionization | ESI |
Ion Mode | Positive and negative |
Description | <Sample processing and extraction> The lyophilized sample in a 2 ml tube was frozen and then homogenized with a 5 mm of zirconia bead by a Mixer Mill (Retsch, Haan, Germany) at 20 Hz for 1 min. Five mg dry weight (DW) of the lyophilized samples were weighed for GC-MS and LC-MS analyses, while 25 mg DW of the samples for CE-MS analysis.
All CE-TOFMS experiments were performed using an Agilent CE capillary electrophoresis system (Agilent Technologies, Waldbronn, Germany), an Agilent G3250AA LC/MSD TOF system (Agilent Technologies, Palo Alto, CA), an Agilent 1100 series binary HPLC pump, and the G1603A Agilent CE-MS adapter and G1607A Agilent CE-ESI-MS sprayer kit. The G2201AA Agilent ChemStation software for CE and the Analyst QS software for TOFMS were used.
Separations were carried out using a fused silica capillary (50 μm i.d. x 100 cm total length) filled with 1 M formic acid for cation analyses or with 20 mM ammonium formate (pH 10.0) for anion analyses as the electrolyte. The capillary temperature was maintained at 20 °C.
The sample solutions were injected at 50 mbar for 15 sec (15 nL). The sample tray was cooled below 4 °C.
Prior to each run the capillary was flushed with electrolyte for 5 min. The applied voltage for separation was set at 30 kV. Fifty percent (v/v) methanol/water containing 0.5 μM reserpine was delivered as the sheath liquid at 10 μL/min.
ESI-TOFMS was conducted in the positive ion mode for cation analyses or in the negative ion mode for anion analyses, and the capillary voltage was set at 4 kV.
A flow rate of heated dry nitrogen gas (heater temperature 300 °C) was maintained at 10 psig.
The fragmentor, skimmer, and Oct RFV voltage were set at 110V, 50V, and 160V for cation analyses or at 120V, 60V, and 220V for anion analyses, respectively.
Automatic recalibration of each acquired spectrum was performed using reference masses of reference standards. The methanol dimer ion ([2M+H]+, m/z = 65.0597) and reserpine ([M+H]+, m/z = 609.2806) for cation analyses or the formic acid dimer ion ([2M-H]-, m/z = 91.0037) and reserpine ([M-H]-, m/z = 607.2661) for anion analyses provided the lock mass for exact mass measurements.
Exact mass data were acqu ired at a rate of 1.5 cycles/sec over a 50-1000 m/z range.
In an every single sequence analysis (maximum 36 samples) on our CE-MS system, we analyzed the standard compound mixture at the first and the end of sample analyses. The detected peak area of standard compound mixture was checked in point of reproducible sensitivity. Standard compound mixture composed of major detectable metabolites including amino acids and organic acids, and this mixture was newly prepared at least once a half year. In all analyses in this study, there were no differences in the sensitivity of standard compounds mixture. |
Comment_of_details |