Sample Set Information

ID SE61
Title LC-MS analysis of parsley
Description Comprehensive analysis of metabolites in parsley (Petroselium crispum, Paramount) was performed by several combination of sample preparation and analytical conditions.

Parsley was hydroponically grown in a greenhouse (S11-S13) or a cultivation shelf (S21-S26) using control medium and stable isotope (15N or 34S)-containing medium.

Authors Sakurai N, Suda K (Kazusa DNA Research Insitute)
Reference Akimoto N, Ara T, Nakajima D, Suda K, Ikeda C, Takahashi S, Muneto R, Yamada M, Suzuki H, Shibata D and Sakurai N (2017) FlavonoidSearch: A system for comprehensive flavonoid annotation by mass spectrometry. Scientific Reports 7: 1243
Comment


Link icon article.png

Sample Information

ID S25
Title Root of cultivation shelf grown parsley (15N labeled)
Organism - Scientific Name Petroselinum crispum, Paramount
Organism - ID NCBI taxonomy:4043
Compound - ID
Compound - Source
Preparation Hydroponically grown in a cultivation shelf. Nitrogen in the culture medium was replaced with 15N. Shoot was harvested.
Sample Preparation Details ID SS2
Comment

Sample Preparation Details Information

ID SS2
Title Cultivation in a cultivation shelf
Description The parsley seeds (Kaneko Seeds Co. LTD., KS100-542, Gunma, Japan) were germinated on a watered filter paper (Whatman 1002-150, GE Healthcare Japan) in a petri dish, and seedlings were grown 14 days in a cultivation shelf under fluorescent lamps (Biolux-A, NEC Corp., Tokyo, Japan) in 16 hr light / 8 hr dark regime at 25 ºC. The seedlings were transplanted on vermiculite (10865595 SK Agri K.K., Gunma, Japan), and grown 164 days under the same condition fertilizing with a culture medium containing 3 mM KNO3, 1 mM Ca(NO3)2, 0.5 mM MgSO4, 0.5 mM NH4H2PO4, 50 µM Fe-EDTA, 49 µM H3BO3, 9 µM MnSO4, 0.8 µM ZnSO4, 0.3 µM CuSO4, and 0.1 µM Na2MoO4. Shoot was harvested, ground to powder in liquid nitrogen by mortar and pestle, and stored at -80 ºC until use.
Comment_of_details

Analytical Method Information

ID M015
Title Method 5: ESI Positive, Data dependent MS2/MS3 scan (FT, IT, IT)
Method Details ID MS5
Sample Amount 5 mg / 20 ul injection
Comment Extractions were triplicated from the same sample and equal volume of the extracts were added and analyzed.

[MassBase ID] MDLC1_48054


Link icon database.png Link icon massbase.png

The raw (binary) and near-raw (text) files of this analysis are available at MassBase.

Analytical Method Details Information

ID MS5
Title LC-FT-MS, ESI, Positive (method 5)
Instrument Agilent1100 HPLC (Agilent), LTQ-FT (Thermo Fisher Scientific)
Instrument Type LC-FTICR-MS
Ionization ESI
Ion Mode Positive
Description Frozen powder of plant material was extracted with three times volumes of methanol containing 25 µM of 7-hydroxy-5-methylflavone as an internal standard. After homogenization using a Mixer Mill MM 300 (Quiagen) at 25 Hz for 2 min twice, homogenates were centrifuged (17,400 g, 5 min, 4 ºC). The supernatant was filtered through 0.2 µm PTFE membrane (Millipore). Hydrophobic compounds in the filtrate were removed by absorbing to C18 silica column (MonoSpin C18, GL Science, Tokyo, Japan).

20 µl of methanol solution was applied to a TSK-gel column ODS-100V (4.6 x 250 mm, 5 µm; TOSOH). Water (HPLC grade; solvent A) and acetonitrile (HPLC grade; solvent B) were used as the mobile phase with 0.1% v/v formic acid added to both solvents. The gradient program was as follows: 3% B (0 min), 97% B (90 min), 97% B (100 min), 3% B (100.1 min), and 3% B (107 min). The flow rate was set to 0.25 ml/min (0-100 min) and 0.5 ml/min (100.1-107 min). The flow rate was set to 0.5 ml/min, and the column oven temperature was set at 40 ºC. Compounds are detected in ESI-positive mode in the range m/z 100-1500. Multistage MSn analyses were carried out using collision induced dissociation (CID) in a linear ion trap detector at a normalized collision energy of 35.0% and an isolation width of 4.0 (m/z), and monitored by both ion trap detector and FT-ICR detector at 25,000 (at m/z 400) mass resolution. The ESI setting was as follows: spray voltage 4.0 kV and capillary temperature 300 ºC. Nitrogen sheath gas and auxiliary gas were set at 40 and 15 arbitrary units, respectively. To monitor HPLC elution, a photodiode array detector was used in the wavelength range 200–650 nm.

Mass scan events were set as follows (referred as method 5): full mass scan with FT-ICR in resolution 12,500, MS2 scans for the most intense 5 ions of full mass scan with IT, and MS3 scans for the most 2 ions of MS2 scan with IT. A dynamic exclusion setting was applied as follows: repeat count, 3; repeat duration, 30 sec; exclusion list size, 500; margin, 10 ppm, and exclusion duration, 20.

The data were acquired by Xcalibur software version 2.07 (Thermo Fisher Scientific).

Comment_of_details [column] TSK-gel ODS-100V (4.6 x 250 mm, 5 micrometer; TOSOH)

[gradient] Solvent A: water containing 0.1% v/v formic acid; Solvent B: acetonitrile containing 0.1% v/v formic acid; Gradient: 3% B (0 min), 97% B (90 min), 97% B (100 min), 3% B (100.1 min), and 3% B (107 min) [total separation time] 107 min