SE207:/MS01
MS Description One hundred microliters of plasma sample w One hundred microliters of plasma sample were extracted with 300 uL of methanol or 1 mL of 75% MeOH (aq). The mixture was mixed and centrifuged. The supernatant was collected for analysis. LC-MS was performed using an Agilent 1200 HPLC system (Agilent Technologies, Santa Clara, CA) coupled to an LTQ Orbitrap XL-MS system (Thermo Fisher Scientific, Inc., San Jose, CA), equipped with an electrospray source operating in either positive- or negative-ion modes. The spray voltage and capillary temperature were 4 kV and 300°C, respectively. The analysis consisted of 2 scan events. Scan event 1 was a full mass type (Analyzer, FTMS; Resolution, 60,000). Scan event 2 was an MS/MS type (analyzer, Ion Trap MS; act type, collision-induced dissociation; normalized collision energy, 35.0%). An aliquot of the extracted sample (5 µL) was injected into a TSK gel ODS-100V reversed-phase column (column size, 3.0 × 50 mm; particle size, 5.0 µm; Tosoh Corp., Tokyo, Japan). The column temperature was set at 40°C. Mobile phases A (0.1% formic acid) and B (acetonitrile with 0.1% formic acid) were used with a gradient of 3% to 97% B from 0 to 15 min, 97% B from 15 to 20 min, 97% to 3% B from 20 to 20.1 min, and 3% B for 4.9 min before the next injection, at a flow rate of 400 µL/min. Data were acquired with Xcalibur software (Thermo Fisher Scientific Inc., San Jose, CA). rmo Fisher Scientific Inc., San Jose, CA).
MS ID MS01  +
MS Instrument Agilent1200 HPLC (Agilent), LTQ Orbitrap XL (Thermo Fisher Scientific)  +
MS Instrument Type LC-Orbitrap-MS  +
MS Ion Mode Positive  +
MS Ionization ESI  +
MS Title LC-Orbitrap, ESI, Positive  +
Modification dateThis property is a special property in this wiki. 18 May 2020 13:40:18  +
hide properties that link here 
  No properties link to this page.
 

 

Enter the name of the page to start browsing from.
Personal tools
View and Edit Metadata
Variants
Views
Actions
Toolbox