SE231:/MS11
MS Comment of details [column] InertSustain AQ-C18 (2.1 x 150 mm [column] InertSustain AQ-C18 (2.1 x 150 mm, 3 micrometer; GL Sciences) [gradient] Solvent A: water containing 0.1% v/v formic acid; Solvent B: acetonitrile; Gradient 2% B (0 min), 2% B (3 min), 98% B (30 min), 98% B (35 min), 2% B (35.01 min), and 2% B (42 min) [total separation time] 42 min B (42 min) [total separation time] 42 min
MS Description - Compound Extraction Approximately 10 m - Compound Extraction Approximately 10 mg of dried fine powder of soybean leaves was accurately weighed in a 2 mL tube, and a 40 times volume of 75% methanol, containing 1 μM of 7-hydroxy-5-methylflavone as an internal standard, was added. Then, the metabolite extraction and the metabolome analysis were performed according to previously described procedures using LC (Nexera X2, Shimadzu), time-of-flight MS (Compact, Bruker), and C18 column (InertSustain AQ-C18, 3 μm x 2.1 mm x 150 mm, GL Sciences) in both ESI positive and negative modes (Sakurai et al., Nucleic Acids Res 51(D1): D660-677, 2023). - Liquid chromatography (LC)-mass spectrometry (MS) analysis LC-MS anaysis was conducted as previously described (Sakurai et al., Nucleic Acids Res 51(D1): D660-677, 2023) using Nexera X2 system (Shimadzu Corporation) and Compact system (Bruker Japan K.K.). An aliquot (2 uL) of the methanol extract was applied to an InertSustain AQ-C18 column (3 um x 2.1 mm x 150 mm, GL Sciences) connected after a guard column (InertSustain AQ-C18 Cartridge Guard Column E, GL Sciences), and separated by water containing 0.1%(v/v) formic acid (Solvent A) and acetonitrile (Solvent B). The gradient program was as follows: 2% B (0 min), 2% B (3 min), 98% B (30 min), 98% B (35 min), 2% B (35.01 min), and 2% B (42 min). The flow rate was set at 0.2 mL/min. The column oven temperature was set to 40 degree C. The compounds separated by the LC were detected using the mass spectrometer under the conditions below: Ionization, Electrospray Ionization (ESI); Polarity, Positive; Scan rate, 1 Hz; Mass scan range, 50-1200; End plate offset, 500 V; Capillary voltage, 4000 V; Nebulizer gas (N2) pressure, 2.5 bar; Dry gas (N2) flow, 8.0 L/min; Dry gas temperature, 200 degree C; Transfer Funnel1 RF, 200.0 Vpp; Funnel2 RF, 200.0 Vpp; In-source CID Energy, 0.0 eV; Hexapole RF, 50.0 Vpp; Quadrupole Ion Energy, 3.0 eV; Low mass m/z, 55.00; Collision energy, 10.0 eV; Collision RF, 450.0 Vpp; Transfer time, 80.0 us; and PrePulse storage, 3.0 us. The data-dependent MS/MS spectra were obtained with the conditions below: Isolation width, 3-15 Da; Collision energy 35 eV; Precursor ion number, 5; Active Exclusion, on; Exclude, after 3 spectra; Release, after 0.2 min; Reconsider precursor, on; and if current intens. / previous intens., 2.0. For mass calibration, 1 mM sodium formate in 50% (v/v) 2-propanol was injected directly into the MS at 38.50–40.50 min of LC separation with a flow rate 0.1 mL/min. The eluent at 0-3 min was wasted. The raw data were obtained by Hystar software (ver.3.2 SR4, Bruker Daltonik, GmbH). ware (ver.3.2 SR4, Bruker Daltonik, GmbH).
MS ID MS11  +
MS Instrument Nexera X2 (Shimadzu), Compact (Bruker Daltonics)  +
MS Instrument Type LC-QTOF-MS  +
MS Ion Mode Negative  +
MS Ionization ESI  +
MS Title LC-Q-Tof-MS, ESI, Negative  +
Modification dateThis property is a special property in this wiki. 11 September 2025 10:08:15  +
hide properties that link here 
  No properties link to this page.
 

 

Enter the name of the page to start browsing from.
Personal tools
View and Edit Metadata
Variants
Views
Actions
Toolbox