SE228:/S040/M01
Sample Set Information
ID | SE228 |
---|---|
Title | LC-MS based untargeted metabolome analysis of various samples |
Description | Compounds in various samples were analyzed using liquid chromatography-mass spectrometry (LC-MS). The same analytical conditions are applied to all samples. Therefore, the compound peaks can be compared to each other by the mass values, the retention time of the LC, and the CID mass spectrum. The data were obtained for the construction of the “Thing Metabolome Repository” website (http://metabolites.in/things). |
Authors | Nozomu Sakurai (National Institute of Genetics, Kazusa DNA Research Institute, email: sakurai (at) kazusa.or.jp) |
Reference | The Thing Metabolome Repository family (XMRs): comparable untargeted metabolome databases for analyzing sample-specific unknown metabolites. Sakurai N, Yamazaki S, Suda K, Hosoki A, Akimoto N, Takahashi H, Shibata D and Aoki Y, Nucleic Acids Research Database Issue) 51 (D1): D660-D677 (2023), DOI: 10.1093/nar/gkac1058 |
Comment |
Sample Information
ID | S040 |
---|---|
Title | Whisky 4 (American) / ウィスキー4(アメリカ) |
Organism - Scientific Name | |
Organism - ID | |
Compound - ID | |
Compound - Source | |
Preparation | The sample was purchased at a local market (2022, Japan). |
Sample Preparation Details ID | SS1 |
Comment |
Sample Preparation Details Information
ID | SS1 |
---|---|
Title | Homogenated and stored at -80 C |
Description | After harvesting, the sample was immediately frozen in liquid nitrogen and stored at -80 degree C. The sample was ground to a fine powder under liquid nitrogen using mortar and pestle and stored at -80 degree C until use. |
Comment_of_details |
Analytical Method Information
ID | M01 |
---|---|
Title | LC-QTOF-MS, ESI, Positive |
Method Details ID | MS01 |
Sample Amount | 0.5 mg / 2 ul injection |
Comment |
Analytical Method Details Information
ID | MS01 |
---|---|
Title | LC-Q-Tof-MS, ESI, Positive |
Instrument | Nexera X2 (Shimadzu), Compact (Bruker Daltonics) |
Instrument Type | LC-QTOF-MS |
Ionization | ESI |
Ion Mode | Positive |
Description | - Compound Extraction
The compounds separated by the LC were detected using the mass spectrometer under the conditions below: Ionization, Electrospray Ionization (ESI); Polarity, Positive; Scan rate, 1 Hz; Mass scan range, 50-1200; End plate offset, 500 V; Capillary voltage, 4000 V; Nebulizer gas (N2) pressure, 2.5 bar; Dry gas (N2) flow, 8.0 L/min; Dry gas temperature, 200 degree C; Transfer Funnel1 RF, 200.0 Vpp; Funnel2 RF, 200.0 Vpp; In-source CID Energy, 0.0 eV; Hexapole RF, 50.0 Vpp; Quadrupole Ion Energy, 3.0 eV; Low mass m/z, 55.00; Collision energy, 10.0 eV; Collision RF, 450.0 Vpp; Transfer time, 80.0 us; and PrePulse storage, 3.0 us. The data-dependent MS/MS spectra were obtained with the conditions below: Isolation width, 3-15 Da; Collision energy 35 eV; Precursor ion number, 5; Active Exclusion, on; Exclude, after 3 spectra; Release, after 0.3 min; Reconsider precursor, on; and if current intens. / previous intens., 2.0. For mass calibration, 1 mM sodium formate in 50% (v/v) 2-propanol was injected directly into the MS at 38.50–40.50 min of LC separation with a flow rate 0.1 mL/min. The eluent at 0-3 min was wasted. The raw data were obtained by Hystar software (ver.3.2 SR4, Bruker Daltonik, GmbH). |
Comment_of_details | [column] InertSustain AQ-C18 (2.1 x 150 mm, 3 micrometer; GL Sciences)
[gradient] Solvent A: water containing 0.1% v/v formic acid; Solvent B: acetonitrile; Gradient 2% B (0 min), 2% B (3 min), 98% B (30 min), 98% B (35 min), 2% B (35.01 min), and 2% B (42 min) [total separation time] 42 min |