MS Comment of details
|
Instrument Type: GC-TOF-MS
|
MS Description
|
Frozen sample was homogenized using mill m … Frozen sample was homogenized using mill mixer with liquid nitrogen. Metabolites were extracted by MeOH/H2O/CHCl3. After liquid‐liquid distribution, upper (MeOH/H2O) layer is used as polar fraction and bottom (CHCl3) layer is used as lipid fraction. Ribitol and nonadecanoic acid methyl ester were used for internal standards of polar and lipid fractions. Polar fraction was evaporated to dryness in a SpeedVac Concentrator. Metabolites were methoximated with methoxyamine hydrochloride/pyridine and then trimethylsilylated with MSTFA..GC conditions: Agilent 6890N GC (Agilent), Carrier gas: He, MS conditions: Pegasus III TOF-MS system (Leco). ditions: Pegasus III TOF-MS system (Leco).
|
MS ID
|
MS2 +
|
MS Instrument
|
6890N GC (Agilent Technologies); Pegasus III TOF-MS system (Leco) +
|
MS Ion Mode
|
Positive +
|
MS Ionization
|
EI +
|
MS Title
|
GC-TOF-MS polar fraction method +
|
Modification dateThis property is a special property in this wiki.
|
22 August 2020 07:06:49 +
|