SE196:/S40/M01
Sample Set Information
ID | SE196 |
---|---|
Title | A lipidome atlas in MS-DIAL 4 |
Description | We formulated mass spectral fragmentations of lipids across 117 lipid subclasses and included ion mobility tandem mass spectrometry (MS/MS) to provide a comprehensive lipidome atlas with retention time, collision cross section and MS/MS information. Using human, murine, algal, and plant biological samples, we annotated and semi-quantified 8,051 lipids by MS-DIAL 4 (http://prime.psc.riken.jp/) with a 1–2% estimated false discovery rate, using the lipidomics standards initiative level 2 or 3 definitions. |
Authors | Hiroshi Tsugawa, Kazutaka Ikeda, Mikiko Takahashi, Aya Satoh, Yoshifumi Mori, Haruki Uchino, Nobuyuki Okahashi, Yutaka Yamada, Ipputa Tada, Paolo Bonini, Yasuhiro Higashi, Yozo Okazaki, Zhiwei Zhou, Zheng-Jiang Zhu, Jeremy Koelmel, Tomas Cajka, Oliver Fiehn, Kazuki Saito, Masanori Arita, and Makoto Arita |
Reference | Tsugawa, H. et al. A lipidome atlas in MS-DIAL 4. Nat Biotechnol (2020). https://doi.org/10.1038/s41587-020-0531-2 |
Comment |
The raw data files are available at DROP Met web site in PRIMe database of RIKEN.
Sample Information
ID | S40 |
---|---|
Title | 40_56_Mouse_Pancreas (Bruker Mouse tissues) |
Organism - Scientific Name | Pancreas (Mouse) |
Organism - ID | NCBI taxonomy 10090 |
Compound - ID | |
Compound - Source | |
Preparation | |
Sample Preparation Details ID | |
Comment |
Analytical Method Information
ID | M01 |
---|---|
Title | 40_Bruker Mouse tissues: Pancreas (Mouse) LC-ESI-MS analysis (DDA) |
Method Details ID | MS05 |
Sample Amount | Positive mode: 1 microLiter; Negative mode: 2 microLiter |
Comment |
Analytical Method Details Information
ID | MS05 |
---|---|
Title | Method 5: Bruker timsTOF Pro DDA for mouse tissue lipidomics |
Instrument | LC, Bruker Elute UHPLC system; MS, Bruker timsTOF Pro |
Instrument Type | UPLC-QTOF-MS |
Ionization | ESI |
Ion Mode | Positive and Negative |
Description | Methanol, isopropanol, and acetonitrile of LC-MS grade were purchased from Wako. Ammonium acetate and EDTA were purchased from Wako and Dojindo, respectively. Milli-Q water was purchased from Millipore.
<LC Method> The LC system consisted of a Bruker Elute UHPLC system. Lipids were separated on an Acquity UPLC Peptide BEH C18 column (50 x 2.1 mm; 1.7 µm) (Waters, Milford, MA, USA). The column was maintained at 45°C at a flow-rate of 0.3 mL/min. The mobile phases consisted of (A) 1:1:3 (v/v/v) acetonitrile:methanol:water with ammonium formate (5 mM) and 10 nM EDTA and (B) 100% isopropanol with ammonium formate (5 mM) and 10 nM EDTA. A sample volume of 0.5-3 µL, which depended on biological samples, was used for the injection. The separation was conducted under the following gradient: 0 min 0% (B); 1 min 0% (B); 5 min 40% (B); 7.5 min 64% (B); 12 min 64% (B); 12.5 min 82.5% (B); 19 min 85% (B); 20 min 95% (B); 20.1 min 0% (B); and 25 min 0% (B). Sample temperature was maintained at 4°C. <MS Method> Mass spectrometric detection of lipids was performed on a hybrid trapped ion mobility-quadrupole time-of-flight mass spectrometer (timsTOF Pro, Bruker Daltonics, Bremen, Germany). Data dependent MS/MS acquisition (DDA) was used. The parameters were MS1 mass ranges, m/z 200-2500 and MS2 mass ranges, m/z 50-2500; MS1 cycle time, 0.5 sec; MS2 accumulation 14 Hz ; collision energy, 30 eV; end plate offset, 500 V; capillary voltage, +4.5 kV/-4.2 kV; nebulizer pressure, 2 bar; dry gas, 10.0 l/min; dry temperature, 220°C/200°C; funnel 1 RF, 300 Vpp; funnel 2 RF, 200 Vpp/250 Vpp; multipole RF, 200 Vpp; deflection delta, +70 V/-70 V; quadrupole ion energy, 6 eV/5 eV; collision transfer energy, 14 eV/15 eV; collision RF, 1100 to 1800 Vpp stepping; transfer time 45 to 75 µs stepping; pre-pulse storage, 10 µs/5 µs; intensity threshold, 31cts.; exclusion time of precursor ion, 0.2 min. The mass calibration was automatically performed using 5 mM sodium acetate calibration solution. |
Comment_of_details |