SE205:/S05/M11
Sample Set Information
ID | SE205 |
---|---|
Title | Untargeted metabolome analysis of plants (6 species) using LC-MS / LC-MSを用いた植物(6種)のノンターゲットメタボローム解析 |
Description | Untargeted metabolome analyses of plants (6 species) using liquid chromatography-mass spectrometry (LC-MS).
The information on the peaks detected in this study was available at the Plant Metabolome Repository website (http://metabolites.in/plants).
検出されたピークは、植物メタボロームレポジトリ (http://metabolites.in/plants) から公開されています。 |
Authors | Sakurai N1,2, Suda K1, Hasebe M3, Ishikawa M3, Kohchi T4, Nishihama R4, Nakai A4, Demura T5, Sano R5, Ohtani M6, Ihara A6, Susa F7, Akimoto N1, Hoshi K1, Osawa S1, Ikeda C1, Ozawa K1, Yamada M1, Muneto R1, Shibata D1 (1 Kazusa DNA Research Institute, 2 National Institute of Genetics, 3 National Institute of Basic Biology, 4 Kyoto Univ., 5 Nara Institute of Science and Technology, 6 Tokyo Univ., 7 RIKEN CSRS), Contact: sakurai AT nig.ac.jp (replace AT with @)
|
Reference | |
Comment |
Sample Information
ID | S05 |
---|---|
Title | Marchantia polymorpha, Takaragaike_1 |
Organism - Scientific Name | Marchantia polymorpha |
Organism - ID | NCBI taxonomy: 3197 |
Compound - ID | |
Compound - Source | |
Preparation | |
Sample Preparation Details ID | |
Comment |
Analytical Method Information
ID | M11 |
---|---|
Title | ESI Negative (Method 1) |
Method Details ID | MS11 |
Sample Amount | 5 mg FW / 20 uL injection |
Comment |
Analytical Method Details Information
ID | MS11 |
---|---|
Title | LC-FT-MS, ESI, Negative |
Instrument | Agilent1100 HPLC (Agilent), LTQ-FT (Thermo Fisher Scientific) |
Instrument Type | LC-FTICR-MS |
Ionization | ESI |
Ion Mode | Negative |
Description | Metabolite extraction
300 mg of powdered sample was mixed with 900 uL of 100% methanol solution containing 25 uM of 7-hydroxy-5-methylflavone as internal standard (IS). The sample was homogenized using Mixer Mill MM 300 (QIAGEN) at 25 Hz for 2 min, twice. The homogenate was centrifuged under 17,400 x g for 5 min at 4 C. A supernatant was passed through a Polytetrafluoroethylene (PTFE) filter (pore size 0.2 um, Millipore), and the filtrate was applied to a C18 silica column (MonoSpin C18, GL Science) to remove highly-hydrophobic contaminants. The filtrate passed through the C18 column was used for LC-MS analyses. The same extraction procedure with 75% methanol was performed without a sample to prepare mock samples and used as negative controls.
|
Comment_of_details | [column] TSK-gel ODS-100V (4.6 x 250 mm, 5 micrometer; TOSOH)
[gradient] Solvent A: water containing 0.1% v/v formic acid; Solvent B: acetonitrile containing 0.1% v/v formic acid; Gradient: 3% B (0 min), 97% B (90 min), 97% B (100 min), 3% B (100.1 min), and 3% B (107 min) [total separation time] 107 min |